Ds500, Ds500+ Operating manual
Original instructions (en)
9. Operation
9.1 Pre-commissioning the Pump
Making the pump ready for use. Ensure the following:
Installation and connection is complete (Section 8.
Pipework is orientated correctly (discharge on top, inlet on
bottom)
Cartridge fitted and locked in place.
Unit connected to mains outlet.
9.2 Starting the Pump For the First
Time
When the pump is first installed, or after a factory reset, the unit
will boot and request the user to:
1. Perform a screen calibration
2. Select a language
The pump will use the following (default) following parameters:
»
Flow rate – 250ml/min
»
Manual mode - On (Ds500+ only)
»
Flow calibration offset ratio – 1.000
»
Auto Restart – Off
»
Fluid Level Monitor – Off
»
Screen Lock - Disabled
»
System PIN code - 1111 (also used for Factory Reset)
»
Wi-Fi - Disabled
Press
to start the pump.
9.3 Stopping the Pump
Press
to stop the pump.
More information on screen calibration, language selection, parameter adjust, etc. can be found in Section 10. Navigating the Screens.