Instructions for Use Evone
56
Parameter
Value / Limits
Units
Measured Parameters
Required time
to change O
2
%
3 s
s
Displayed Values
All pressures and volumes are calculated into ATPD:
ambient temperature (variable) and ambient
pressure (variable), dry (no humidity). In the lungs,
a temperature of 36 °C and 100% humidity holds.
-
Pressure Lumen Accuracy
+/- 0.5% FSS = +/- 1 mbar
%
Oxygen Measurement
(compensated for
ambient pressure)
0 – 100 (compensated for ambient pressure)
%
Sound pressure level
52
dBA
Environmental Parameters
Water Resistance
IP 20
-
Operation
Temperature
10 – 35
°C
Barometric Pressure
700 – 1060
HPa
Relative Humidity
30 – 75%, non-condensing
RH
Storage
Relative Humidity
10 – 95%, non-condensing
RH
Temperature
-15 – 40
°C
Barometric Pressure
500–1060
HPa
Stacking
Do not stack
-
CO
2
Measurement Specifications
Principle of Operation
Non-dispersive infrared, single beam optics, dual
wavelength, no moving parts.
-
Initialization Time
Less than 10 seconds at an ambient temperature of
25 °Celsius, full specifications within 1 minute.
-
CO
2
Measurement Range
0 – 70
mmHg
CO
2
Resolution
0.1
mmHg
CO
2
Accuracy
± 2 or ± 5% whichever is higher
mmHg
CO
2
Stability
< 0.5
mmHg
per hour
CO
2
Noise
< 0.25 at 5% CO
2
mmHg
CO
2
sampling rate
100
Hz
Gas standard
CO
2
values are displayed in BTPD
Note 1: EtCO
2
values are determined as the end value of the CO
2
curve during expiratory phase.
Note 2: Different frequencies and immunity levels are not expected in the intended environment.