Technical
Specifications
55
10
Parameter
Value / Limits
Units
Standard
Emissions class/group or Immunity test level
IEC 61000-4-3
RF wireless communications equipment:
Frequency: Immunity test level:
930 MHz
28 V/m
1720 MHz
28 V/m
1845 MHz
28 V/m
1970 MHz
28 V/m
2450 MHz
28 V/m
5240 MHz
9 V/m
5500 MHz
9 V/m
5785 MHz
9 V/m
-
IEC 61000-4-4
AC power:
± 2 kV, 100 kHz repetition frequency
Serial Interface: ± 1 kV, 100 kHz repetition frequency
-
IEC 61000-4-5
AC power (line-to-line):
± 1 kV
AC power (line-to-ground): ± 2 kV
-
IEC 61000-4-6
3 V
rms
, 0,15 MHz – 80 MHz
6 V
rms
in ISM bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
-
EN 61000-4-8
30 A/m
-
IEC 61000-4-11
0% UT; 1 cycle
70% UT; 25 cycles
>5% UT; 250 cycles
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
-
Gas Supply Parameters
Gas Supply Pressure
3.0 – 6.0 +/- 10%
bar
Gas Supply Flow
20
L / min
Gas Supply Types
Medical grade oxygen and medical grade
compressed air
-
Ventilation Parameters
Inspiration Flow
2 – 20
L / min
Inspiration Flow Accuracy
4% of setpoint or 0.1 %FS (FS = 100l/min)
whichever is greater
%
FiO
2
21 – 100
%
I:E ratio /
inspiration percentage
FCV
®
1:1.0 – 1:2.5 / Jet 20–50%
-
Driving Pressure
0.1 – 1.5
bar
Frequency
60 – 150
BPM
Pressure Set Range
-10 – 100 ± 0.5
mbar