Configur
ation
37
6
The (alarm) limits can be adapted manually before starting and during ventilation as indicated
in Table 6.3.
Parameter
Unit
Setting
Resolution
Default
Min
(Rounded)
Max
(Rounded)
(Rounded)
Peak alarm
mbar
10
105
1
25
kPa
1.0
10.5
0.1
2.5
cmH
2
O
10
107
1
25
EEP alarm
mbar
-15
25
1
-2
kPa
-1.5
2.5
0.1
-0.2
cmH
2
O
-15
25
1
-2
EtCO
2
upper alarm
mmHg
20
70
1
50
kPa
2.7
9.3
0.1
6.7
vol%
2.6
9.2
0.1
6.6
EtCO
2
lower alarm
mmHg
5
50
1
26
kPa
1.0
6.7
0.1
3.4
vol%
1.0
6.6
0.1
3.4
Maximum
Inspiratory Volume
mL
150 – 1500
50
10 mL/kg IBW
Minimum
Inspiratory Volume
mL
150 – 1000
50
4 mL/kg IBW
Table 6.3 Setting ranges and default values of (alarm) limits in FCV
®
mode
After adjusting ventilation settings of alarm limits, changes need to be confirmed.
FCV
®
is initiated by pushing the start button. Evone will start with
1 volume controlled ventilation cycle to:
• confirm equal pressure values while independently measured via the ventilation lumen
and via the pressure measurement lumen
• achieve the desired inspiratory volume in a graduate and safe way
• detect leakage
Warning: Do not set the volume limits too wide from each other. Risk on barotrauma.
It is recommended to set limits about 150 mL upper and lower of the expected
tidal volume.