PART 3 - Technical manual
One of the bars will only move in
one direction
Is the plug properly connected to the control unit and the actua-
tor?
Is the cable intact?
Switch the plugs of the two bars. If the same bar still does not
work, then the bar motor is defective. If the other bar is not
working, then the hand control or the control unit is defective.
eplace the hand control. If this does not solve the problem,
replace the control unit.
Reset the system by simultaneously pressing the two top buttons
on the hand control. You will hear a series of rhythmic beeps.
Press and hold the buttons until the beeping rhythm changes.
Replace the bar motor.
Contact a qualified technician or Healthcare Lifting Specialist.
The hand control buttons are not
working
Is the plug properly connected to the control unit?
Is the cable intact?
If the actuator still does not work, try changing the channel.
Replace the actuator if this does not solve the problem.
Reset the system by simultaneously pressing the two top buttons
on the hand control. You will hear a series of rhythmic beeps.
Press and hold the buttons until the beeping rhythm changes.
Replace the hand control.
Contact a qualified technician or Healthcare Lifting Specialist.
One or more actuators will not go
up or down
Is the plug properly connected to the control unit?
Is the cable intact?
If the actuator still does not work, try changing the channel.
Replace the actuator if this does not solve the problem.
Reset the system by simultaneously pressing the two top buttons
on the hand control. You will hear a series of rhythmic beeps.
Press and hold the buttons until the beeping rhythm changes.
Contact a qualified technician or Healthcare Lifting Specialist.
One or more actuators / actuator
fittings will not move entirely
down
Are the actuator fittings perpendicular to the bed frame?
Have the side rail dimensioning requirements been met, cf. page
30? If not, the free movement of the actuator fittings may be
affected.
Dirt, dust etc. may obstruct the sliding function of the actuator
fitting. Clean the actuator fitting and lubricate it with silicone
spray.