35
PART 3 - Technical manual
Mounting the actuator fittings with actuator
The four actuator fitting cables are colour-
coded.
Actuator fittings with green actuator plug O
rings are mounted to the right-hand side of the
patient. Actuator fittings with yellow actuator
plug O rings are mounted to the left-hand side
of the patient.
1. The actuator fitting and actuator is
inserted into the mounting bracket from
the bottom. Push the red button and pull
out the adjustment screw to guide the
actuator fitting to the desired position.
2. Release the adjustment screw and
allow it to fall into place to fix the
actuator fitting in the desired position/
at the desired height. Apply pressure to
the adjustment screw to ensure its full
insertion. Apply pressure to the actuator
fitting to check if it has locked in place.
3. Use this procedure to attach all four
actuator fittings
Mounting the support bars
Mount the support bar on each side between
the two actuator fittings. You will need two 8 x
20 mm button head bolts and two self-locking
M8 nuts with M8 washers for the task.
Attach the support bar to the top hole of the
angle bracket on the actuator fitting.
The flat side of the support bar should be
facing the mattress.
Insert the actuator fitting in the mounting
bracket
Pull out the adjustment screw
Mounting the support bar
The flat side of the support bar should be
facing the mattress