VEGASON 51K … 53K
41
Set-up
Hinweis:
When connecting the adjustment software
(VVO) to a sensor from which data has al-
ready been saved, you are asked if the
saved data should be transferred to the
sensor or if you want to transfer the sensor
data to the database of VVO (and overwrite).
If you get no sensor connection, check the
following:
- is the sensor powered with supply voltage
(min. 19 V)
!
- when VEGACONNECT 2 is connected to
the signal line, is the load resistor 250 …
350
W!
- did you erroneously use a
VEGACONNECT instead of the new VEGA-
CONNECT 2!
- have you connected VEGACONNECT 2 to
COM1 on the PC?
Configuration
• Choose the menu "Configuration/Measure-
ment loop“ to get further information on the
sensor type, the software version of the
sensor, the meas. unit, the measurement
loop designation etc.
• Click on "Quit“.
• Click on the menu "Configuration/Measure-
ment loop/Modify“. This is the first step to
set-up the sensor.
In the menu "Modify meas. loop configuration“
you can give a name to the measurement
loop (e.g. vessel 10) and co-ordinate a
measurement loop description (e.g. sludge
separator).
• Now enter in this menu whether a level, a
distance or a gauge should be measured
and click on "OK“.