4.2 Measurement of liquids
Flat vessel top
On flat vessels the mounting is mainly made
on a very short DIN-socket piece. Reference
plane on flange versions is the instrument
flange. The transducer should protrude out of
the flange pipe.
Flange version on very short DIN-socket piece
Ideal is the mounting directly on the vessel
top. A round opening on the vessel is suffi-
cient to fasten the VEGASON 53 sensor with
a compression flange.
Flange version (compression flange) on flat vessel
top
Reference plane
Reference plane
Min.
distance
£
60 mm
0
0 m
7 m
50 %
1,2
1,2
m
0,4
0,4
100 %
5,5˚
12˚
4 m
Meas. distance
0
0 m
15 m
100 %
50 %
1,2
1,2
m
0,4
0,4
3˚
8˚
Meas. distance
Mounting and installation
VEGASON 51/52
VEGASON 53
emitted power
emitted power
emitted power
emitted power