1.
Highlight a desired timeslot number with the cursor.
2.
Press ‘Enter’ to invert current status of the timeslot. Any timeslot marked with “*” symbol will be included in
measured object.
3.
Press ‘F1’ to select all the time slots.
4.
Timeslots that is not valid for selection as a measured object will be highlighted with red color.
5.
Press ‘F1’ to make both Receive and Transmit sections the same.
6.
When measured object is selected press ‘F2’ to apply new settings. To move back to previous menu
without saving new object parameters, press ‘ESC.’
Protective resistors
The Protective resistor field provides the control for built-in protective resistors (470Ohms). These resistors are
used only in “Monitor” and “Transit-Monitor” operational modes.
If the field is set to “Off” value protective resistors are shunted.
Test pattern
This menu allows the user to set test pattern parameters and provides control for the test pattern sending. The
“Test pattern” screen is shown at the Figure
5.4
below.
Send pattern
The Send field allows to start/stop the insertion of current test pattern into timeslots selected in the “Line
interface/Test speed” menu.
Type
The Type field allows the user to select test pattern type from the following list:
All
1: All1
All 0
: All 0
55
: 55 Octet
2
е
6
: Pseudo-random sequence
2
е
9
: Pseudo-random sequence
2
е
11
: Pseudo-random sequence
2
е
15
: Pseudo-random sequence
2
е
23
: Pseudo-random sequence
User
: User-defined sequence
Byte 1, Byte 2, Byte 3
Figure 5.4:
“Test pattern” menu
Page 15 of 30
TX50e e-Manual D07-00-009 REVA00