Electric Force (EFM) Imaging
Electric Field Gradient Detection—Preparation
252
MultiMode SPM Instruction Manual
Rev. B
Voltage Applied to the Tip
Notice that the jumper con
fi
connects the Analog 2 signal from the
NanoScope III controller (± 12 VDC range) to the tip, and is
exactly the same
as the jumper
con
fi
guration shown in
fi
guration as shipped from the factory.
Note:
In all con
fi
gurations which apply voltage to the tip, an E-
fi
eld cantilever holder
is required. Contact Veeco for more information.
Enabling the Analog 2 Voltage Line
The “Analog 2” voltage line is normally used by the NanoScope to control the attenuation (1x or
8x) of the main feedback signal. This application of EFM imaging uses the Analog 2 signal for
EFM data. Therefore, input attenuation must be disabled for the duration of the EFM experiments.
To do this, select
di
>
Microscope Select
>
Edit
>
Advanced
to display the
Equipment
dialog
box. Set
Analog 2
to
User-de
fi
ned
. Return to the main
Feedback Controls
panel; the
Analog 2
fi
eld should now be enabled. This signi
fi
es that voltage is now being supplied via the “Analog 2”
pin located on the MultiMode baseplate header.
Remember to restore (
Allow in attenuation
) upon completion of EFM imaging.
Figure 14.3h
Jumper con
fi
guration for application of voltage to tip
(for systems
with
the Basic Extender Module).
Sample
Tip
Analog 2
Ground
Analog 2
Unused
Auxiliary D (to NanoScope III controller)
To AFM Tip
Indicates jumpers
Analog 2
STM
Piezo Cap