Magnetic Force (MFM) Imaging
Software Setup Con
fi
guration (Basic, Quadrex or NSIV)
234
MultiMode SPM Instruction Manual
Rev. B
13.4 Software Setup Con
fi
guration (Basic, Quadrex or NSIV)
1. Select
di
>
Microscope Select
to display the
Microscope Select
dialog box (see
Figure 13.4a
Microscope Select Dialog Box
2. Click the
Edit
button to open the
Equipment
dialog box.
3. Select the appropriate
Controller
.
4. Select the appropriate
Extender
.
Note:
This step is not necessary for NanoScope IV.
5. Click the
Ok
button when complete.
13.5 Troubleshooting Suggestions
13.5.1 MFM Image Veri
fi
cation
The procedure described above should produce a good magnetic force gradient image of the
videotape sample. If there is a problem, check that the
Interleave Mode
is set to
Lift
, that
Interleave
is
Enabled
and that the
Scan Line
is set to
Interleave
. Check also that the
Interleave
values of
Drive Amplitude
and
Drive Frequency
are initially set equal to the main
Scan Controls
values.
012