Disassembly of the Flow Cell
If the contamination in the flow cell cannot be removed by the
aforementioned cleaning procedure, disassemble the cell and
clean its parts as instructed below.
Figure 21 Flow Cell Unit
1. Loosen setscrews A and detach the shielding plate.
2. Detach the cell window.
3. Loosen the upper and lower retaining screws 1 and 2, see
Figure 22.
4. Detach the cell clamp.
5. Take out the flow cell.
Setscrews A
Shielding plate