03-914875-00:1
Figure 12 Removal of Light Source Cover
WA R N IN G :
E XP LO SIO N H A ZA R D
•
Explosion of the xenon lamp could
cause serious injury.
High-pressure gas at about 1 MPa is
filled in the lamp at room temperature.
•
Be careful not to apply force to the
lamp bulb (glass part) when replacing
the lamp.
•
Wear protective goggles and gloves
when handling the lamp.
2. When placing the xenon lamp in the lamp holder,
(Figure 13), put the anode terminal (marked '+') of the
lamp into the metal fixture on the ceramic plate. Then,
using the anode nut, mount the lamp so that the
protruding part of the bulb is positioned as shown in
Figure 14. Do not touch the quartz envelope while the
lamp is being installed. Handle the lamp by holding the
metal pieces on each end of the lamp.
A thin metallic wire is attached on the center area of the
bulb, as shown in Figure 14. This wire serves to ensure
that the lamp is easily started. When handling the lamp,
be careful not to break this wire. If the thin metallic wire
is not located on the side of the lamp with the small
Mounting
Screws
Light source
cover
Mounting
Screws