Vante
®
Model 4600
12
WARNING
: Always operate the Sealing Head with the splash guard in place. Do NOT
pull the tubing through the sealing region. Under
no
circumstances should
the tubing be pulled at the instant of sealing. This may cause an opening in
the tubing, which will subject the media to non-sterile conditions and the
user to potentially hazardous fluids. In addition, the Sealing Head may
become damaged by arcing caused by contamination of the sealing region.
WARNING: Do not place fingers in close proximity to the sealing region while sealing.
Doing so will subject the operator to a radio frequency burn hazard.
Figure 2.8 Incorrect Sealing Method
The seal is completed in approximately one second, as indicated when the sealing jaw
retracts. Another seal cannot be made until the sealing jaw fully retracts and the seal
trigger has been completely released.
To make another seal, lift the tubing up to release the trigger and push the tubing to the
next seal position (see Figure 2.9). Activate the seal trigger with the tubing to repeat the
process. The completed seals may be moved to either the right or left, but for best results,
move the completed seals to the left (while facing the head shell) to avoid their
interfering with the seal trigger. For more information, refer to the instruction manual for
the Model 4605 Sealing Head.
Содержание 4600
Страница 1: ...Doc No 46000810 01 Rev D INSTRUCTION MANUAL MODEL 4600 RF Tube Sealer Power Source...
Страница 2: ...ii This page intentionally left blank...
Страница 4: ...iv This page intentionally left blank...
Страница 8: ...viii This page intentionally left blank...
Страница 12: ......
Страница 32: ...Vante Model 4600 20 This page intentionally left blank...