11
WARNING:
If you observe arcing while making a seal, follow the instructions in
the Troubleshooting Guide section under “Arcing or Bad Seals.”
Although a seal made when you experience arcing may look
acceptable, it may be inadequate. Take precautions as if this is an
inadequate seal.
WARNING
:
Under
no
circumstances should the tubing be separated by pulling it
at the instant of sealing. This may cause an opening in the tubing,
which will subject the media to non-sterile conditions and the user to
potentially hazardous fluids. In addition, the sealing head may
become damaged by arcing caused by contamination of the sealing
region.
2.3
Sealing Procedure with Model 4605 Bench Top Sealing Head
NOTE:
For clarity, the splash guard is not shown in some of the figures in this
manual; however, it should always be in place during the sealing process.
To make a seal, lay the tubing between the jaws (“sealing region”), allowing the weight
of the tubing to press against the seal trigger (a slight additional manual force may be
required), as shown in Figure 2.7. When the trigger is activated, the RF sealing jaw will
compress the tubing against the ground jaw and RF power will be applied for
approximately one second. The front plate of the Sealing Head will glow during RF
application. After the seal is formed, the sealing jaw will retract to its original position.
Figure 2.7 Correct Sealing Method
Содержание 4600
Страница 1: ...Doc No 46000810 01 Rev D INSTRUCTION MANUAL MODEL 4600 RF Tube Sealer Power Source...
Страница 2: ...ii This page intentionally left blank...
Страница 4: ...iv This page intentionally left blank...
Страница 8: ...viii This page intentionally left blank...
Страница 12: ......
Страница 32: ...Vante Model 4600 20 This page intentionally left blank...