Vante
®
Model 4600
14
2.4
Seal Spacing
The Sealer is designed to allow repetitive seals to be made on a length of tubing filled
with media. However, several factors control how closely the seals may be spaced.
Seals which progress along an open-ended length of tubing may be spaced as desired, but
½ inch (1 cm) is the recommended minimum.
Seals which progress along a
close-ended
length of tubing, as in a segmenting process,
must be spaced
no less
than the bag manufacturer’s recommendation, to avoid rupture of
the tubing due to pressure build-up.
Multiple seals in close proximity are not required, nor are they recommended when using
the Power Source. If multiple seals must be used, they should be spaced at least one inch
(2.5 cm) apart, as shown in Figure 2.10.
Figure 2.10 Multiple Seal Spacing
WARNING
:
Never attempt to make segmentation seals closer than recommended
without validating for seal integrity.