Vante
®
Model 4600
10
WARNING
:
Dimming of the indicator light will occur, but do not release the lever
until
after
the indicator light is completely off! Premature lever
release will cause incomplete sealing and/or ruptured tubing. Holding
the lever closed will
not
cause overheating or burn-through of the
tubing.
CAUTION:
Never squeeze the lever when non RF-reactive materials are in the sealing
region or if the Power Source is not connected and turned on. Damage to
the lever and poor Power Source operation may result.
The seal is typically completed in one second. The lever should be held closed for two
seconds after the sealing head light completely extinguishes. At that time, the lever may
be released.
Another seal cannot be made until the moveable lever has been fully opened.
See Figure 2.6. For more information refer to the Instruction Manual for the Model 4105
Sealing Head.
Figure 2.6 Releasing the Sealing Head Lever at Seal Completion
NOTE:
The Power Source is equipped with a thermal protection switch to prevent
the equipment from becoming damaged. If thermal cutout is reached, the
power source will not work for several minutes until the temperature of
the equipment lowers.
WARNING
:
The Power Source is designed for rapid sealing applications;
however, extreme heavy duty use may result in thermal build up in
the jaw area which can cause poor seal quality or tubing ruptures. If
you should experience these problems, reduce the rate of sealing, as
necessary, or allow for periods of rest for the Sealing Head.
Содержание 4600
Страница 1: ...Doc No 46000810 01 Rev D INSTRUCTION MANUAL MODEL 4600 RF Tube Sealer Power Source...
Страница 2: ...ii This page intentionally left blank...
Страница 4: ...iv This page intentionally left blank...
Страница 8: ...viii This page intentionally left blank...
Страница 12: ......
Страница 32: ...Vante Model 4600 20 This page intentionally left blank...