IRM-5000P USER’S MANUAL
REV 2
50
4.3
Setting the Temperature Units of Measure
When performing tests, you can provide the Device Under Test (D.U.T.) temperature and the
IRM-5000P will automatically calculate the equivalent resistance values at 40
°
C and 20
°
C. You
can select the preferred units of measure for entering the D.U.T. temperature by using the
steps below:
a.
Start from the “START-UP” menu:
Press the
[2]
key (
UTILITIES
).
b.
The following screen will be displayed:
Press the
[3]
key (
SETUP
).
c.
The following screen will be displayed:
Press the
[5]
key (
°
C/
°
F
)
.
d.
The following screen will be displayed:
Press the
[1]
key (
°
C
) or the
[2]
key (
°
F
) to select the corresponding units of measure.
The D.U.T. temperature units of measure will be set and you will be returned to the
“START-UP” menu.
1.
°
C
2.
°
F
TERMINAL VTG: 1 V
1.TEST ID 2.CLOCK
3.COMP INT 4.50/60Hz
5.
°
C/
°
F
TERMINAL VTG: 1 V
1.PRINT 2.SAVE/RES
3.SETUP
TERMINAL VTG: 1 V
1.
RUN TEST 05/05/10
2.
UTILITIES 15:15:15
Bat:
TERMINAL VTG: 1 V
Содержание IRM-5000P
Страница 25: ...REV 2 IRM 5000P USER S MANUAL 21 Figure 6 Current Plot from a Resistance Test...
Страница 36: ...IRM 5000P USER S MANUAL REV 2 32 Figure 10 Step Voltage Test Results Report Printout...
Страница 37: ...REV 2 IRM 5000P USER S MANUAL 33 Figure 11 Step Voltage Test Resistance and Current Plots...
Страница 42: ...IRM 5000P USER S MANUAL REV 2 38 Figure 13 Dielectric Discharge Test Resistance Plot...
Страница 43: ...REV 2 IRM 5000P USER S MANUAL 39 Figure 14 Dielectric Discharge Test Current Plot...
Страница 49: ...REV 2 IRM 5000P USER S MANUAL 45 Figure 15 Typical Test Record Directory Printout...