REV 2
IRM-5000P USER’S MANUAL
11
3.0 OPERATING
PROCEDURES
3.1 Performing
Tests
3.1.1. Entering Test Record Header Information
You can enter the test record header information before performing tests. The record header
includes identifying information such as company, location, test description, etc. Once the
header information has been entered, it will apply to all subsequent test records. To enter the
header information:
a.
When the unit is turned on, you will be presented with the “START-UP” menu:
NOTES
•
If the IRM-5000P is running on battery power, the battery icon will be
displayed on the LCD screen as shown above. The icon also represents the
charge level of the battery.
If the IRM-5000P is plugged in, a plug icon will be displayed on the LCD
screen as shown below:
•
The test terminal voltage is displayed at the bottom of the LCD screen. The
IRM-5000P can be used as a voltmeter by connecting the positive and
negative leads to a voltage source. The voltage measurement range is
from 50-1250 V(AC/DC).
Press the
[2]
key (
UTILITIES
).
b.
The following screen will be displayed:
Press the
[3]
key (
SETUP
).
1.PRINT 2.SAVE/RES
3.SETUP
TERMINAL VTG: 1 V
1.
RUN TEST 05/05/10
2.
UTILITIES 10:10:15
Bat:
TERMINAL VTG: 1 V
1.
RUN TEST 05/05/10
2.
UTILITIES 10:10:15
Bat:
TERMINAL VTG: 1 V
Содержание IRM-5000P
Страница 25: ...REV 2 IRM 5000P USER S MANUAL 21 Figure 6 Current Plot from a Resistance Test...
Страница 36: ...IRM 5000P USER S MANUAL REV 2 32 Figure 10 Step Voltage Test Results Report Printout...
Страница 37: ...REV 2 IRM 5000P USER S MANUAL 33 Figure 11 Step Voltage Test Resistance and Current Plots...
Страница 42: ...IRM 5000P USER S MANUAL REV 2 38 Figure 13 Dielectric Discharge Test Resistance Plot...
Страница 43: ...REV 2 IRM 5000P USER S MANUAL 39 Figure 14 Dielectric Discharge Test Current Plot...
Страница 49: ...REV 2 IRM 5000P USER S MANUAL 45 Figure 15 Typical Test Record Directory Printout...