![Vanguard Instruments Company IRM-5000P Скачать руководство пользователя страница 12](http://html2.mh-extra.com/html/vanguard-instruments-company/irm-5000p/irm-5000p_user-manual_3393196012.webp)
IRM-5000P USER’S MANUAL
REV 2
8
2.0 PRE-TEST SETUP
2.1 Operating
Voltages
The IRM-5000P’s operating voltage is preset at the factory and is selectable between 100-120
Vac, 50/60 Hz or 200-240 Vac, 50/60 Hz. The voltage is set by placing jumpers at the JP16 and J1
jumper blocks on the main board as listed in Table 3 and illustrated in Figure 2 and Figure 3.
Table 3. Voltage Selection Jumper Settings
Voltage
Selection
JP16 Settings
J1 Settings
100 – 120 Vac
Jumper Pin 1 and 2
Jumper Pin 3 and 4
Jumper Pin 1 and 2
Jumper Pin 3 and 4
200 – 240 Vac
Jumper Pin 2 and 3
Jumper Pin 2 and 3
Figure 2. 100-120 Vac Jumper Settings
Figure 3. 200-240 Vac Jumper Settings
Содержание IRM-5000P
Страница 25: ...REV 2 IRM 5000P USER S MANUAL 21 Figure 6 Current Plot from a Resistance Test...
Страница 36: ...IRM 5000P USER S MANUAL REV 2 32 Figure 10 Step Voltage Test Results Report Printout...
Страница 37: ...REV 2 IRM 5000P USER S MANUAL 33 Figure 11 Step Voltage Test Resistance and Current Plots...
Страница 42: ...IRM 5000P USER S MANUAL REV 2 38 Figure 13 Dielectric Discharge Test Resistance Plot...
Страница 43: ...REV 2 IRM 5000P USER S MANUAL 39 Figure 14 Dielectric Discharge Test Current Plot...
Страница 49: ...REV 2 IRM 5000P USER S MANUAL 45 Figure 15 Typical Test Record Directory Printout...