IRM-5000P USER’S MANUAL
REV 2
30
g.
The following screen will be displayed while the test is being performed:
The LCD screen will be dynamically updated displaying the remaining time along with
the resistance, current, and test voltage values. When the test is completed, the results
for the first time interval and voltage step will be displayed as shown:
Press any key to continue. The results for the second time interval and voltage step will
be displayed:
Press any key to continue. The results for the third time interval and voltage step will be
displayed:
Press any key to continue. The results for the fourth time interval and voltage step will
be displayed:
Press any key to continue. The results for the final time interval and voltage step will be
displayed:
Press any key to continue.
h.
The following screen will be displayed:
9.863G@ 05:00, 2500V
2500V 0.00uF 05:00
TERMINAL VTG: 1 V
9.847G@ 04:00, 2000V
2500V 0.00uF 05:00
TERMINAL VTG: 1 V
9.851G@ 03:00, 1503V
2500V 0.00uF 05:00
TERMINAL VTG: 1 V
9.973G@ 02:00, 1002V
2500V 0.00uF 05:00
TERMINAL VTG: 1 V
9.952G@ 01:00, 499V
2500V 0.00uF 05:00
TERMINAL VTG: 1 V
Bat:
Time: 04:59
Resistance: 10.002 G
Current: 50.2 nA
TEST VOLTAGE: 498 V
Содержание IRM-5000P
Страница 25: ...REV 2 IRM 5000P USER S MANUAL 21 Figure 6 Current Plot from a Resistance Test...
Страница 36: ...IRM 5000P USER S MANUAL REV 2 32 Figure 10 Step Voltage Test Results Report Printout...
Страница 37: ...REV 2 IRM 5000P USER S MANUAL 33 Figure 11 Step Voltage Test Resistance and Current Plots...
Страница 42: ...IRM 5000P USER S MANUAL REV 2 38 Figure 13 Dielectric Discharge Test Resistance Plot...
Страница 43: ...REV 2 IRM 5000P USER S MANUAL 39 Figure 14 Dielectric Discharge Test Current Plot...
Страница 49: ...REV 2 IRM 5000P USER S MANUAL 45 Figure 15 Typical Test Record Directory Printout...