![Vaisala MAWS101 Скачать руководство пользователя страница 120](http://html1.mh-extra.com/html/vaisala/maws101/maws101_user-manual_3205925120.webp)
User's Guide _______________________________________________________________________
118 __________________________________________________________________ M210243en-A
0201-046
Figure 107
Selecting an Upload Configuration File
When the file has been transferred, MAWS starts logging data
according to the settings in the configuration file.
After a system update (for instance, adding new sensors to your
MAWS), the configuration file needs to be updated to match the
changes. Normally, you need to modify the configuration file yourself
with the MAWS Lizard setup software.
CAUTION
When you upload a new configuration, the system erases all data log
files from MAWS. Make sure you download the files you want to
save before uploading the new configuration. For more information
on downloading data log files, see section Selecting Files for
Downloading on page 122.
Data Logging
Logging means storing of the measured and calculated data in the
MAWS internal memory, a 2 MB Flash chip. Logged data can later be
retrieved via serial line.
Logged data is stored in daily files, for example, L2010326.dat that is
a binary file. Naming convention is explained below:
- All log files begin with the name of the log group (max. two
characters) which is followed by the date.