Chapter 6 ___________________________________________________Calibration and Adjustment
VAISALA _______________________________________________________________________ 57
Using MI70
To adjust temperature:
1.
Connect MI70 to MMT162 via connector II.
2.
Turn on MI70.
3.
On the MI70 display, go to
Menu
→
Functions
→
Adjustments
.
4.
Follow the instructions given on the MI70 display.
5. Select
T
from the list and press
Select.