One-Point Temperature Adjustment
The MMT162 must be unpowered during stabilization which can take 30
minutes or more.
Using Serial Line
1.
Remove the probe filter and insert the probe into the reference
temperature.
2.
Enter the command
CT
and press
ENTER
.
CT
3. Type
C
and press
ENTER
a few times to check if the reading is
stabilized. Let the reading stabilize, give the reference temperature
after the question mark and press
ENTER
three times
.
Example
:
ct
T : 22.03 Ref1 ?
T : 22.03 Ref1 ?
T : 22.03 Ref1 ? 25.0
Press any key when ready ...
T : 22.02 Ref2 ?
OK
4.
OK
indicates that the calibration has succeeded.
5.
Take the probe out of the reference conditions and replace the
filter.