Configuring Serial Line Operation
Set Serial Line Settings
Use the serial line command
SERI
[
b p d s
] to set communication
settings for the user port.
SERI
[
b p d s
]
where
b
=
Bit rate (110, 150, 300, 600, 1200, 2400, 4800, 9600,19200,
38400)
p
=
Parity (n = none, e = even, o = odd)
d
=
Data bits (7 or 8)
s
=
Stop bits (1 or 2)
Example:
>SERI 600 N 8 1
600 N 8 1
>
You need to reset the transmitter to activate the new communication
settings set with the
SERI
command.
The settings can be changed one parameter at a time or all parameters at
once:
>SERI O
changing parity only
4800 O 7 1
>SERI 600 N 8 1
changing all parameters
600 N 8 1
>
You can use the
SERI
command to change/view the user port settings
even if you are currently connected to the service port.