User's Guide ______________________________________________________________________
34 __________________________________________________________________ M210934EN-B
Calibrate Analog Output
Use the
ACAL
command to adjust analog output channels. To adjust
channel 1, use parameter 1 and to adjust channel 2, use parameter 2 as in
the example below.
After entering the
ACAL
command, use a calibrated current/voltage
meter to measure the output, and enter the values.
Example
(current outputs):
acal 1
Ch1 U1 ( V ) ? 1.001
Ch1 U2 ( V ) ? 9.011
acal 2
Ch2 U1 ( V ) ? 0.0988
Ch2 U2 ( V ) ? 0.8997
Set Malfunction Alarm Output for Analog
Channels
Use the
AERR
command to set the analog output levels that are output in
case of a device malfunction. The default output level is 0.
AERR
Example:
aerr
Ch1 error out : 10.000V ? 0
Ch2 error out : 1.000V ? 0
NOTE
The error output value must be within a valid range of the output
mode.
NOTE
The error output value is displayed only when there are minor
electrical faults such as a humidity sensor damage. When there is a
severe device malfunction, the error output value is not necessarily
shown.