![Vaisala HUMICAP MMT162 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/vaisala/humicap-mmt162/humicap-mmt162_user-manual_3376211018.webp)
User's Guide ______________________________________________________________________
16 __________________________________________________________________ M210934EN-B
0805-006
Figure 4
Installing the Transmitter
4.
Use a 30 mm wrench to tighten the connection to 25 Nm, as shown
in Figure 4 above. If you do not have a 30 mm wrench, use a
1 3/16” wrench or an adjustable wrench instead.
CAUTION
Only tighten the probe from the 30 mm nut. Do NOT apply force to
other points in the probe body.
5.
Connect the wires of the connection cable.
0710-130
Figure 5
Connecting the Cable
6.
Plug in the cable to the transmitter. Be sure to use the correct
connector; refer to section Wiring on page 17.
7.
Cover the unused connector on the transmitter with the rubber plug
that is attached to the transmitter.