Chapter 6 ___________________________________________________Calibration and Adjustment
VAISALA _______________________________________________________________________ 53
6.
When the reading is stable, give the reference humidity after the
question mark and press
ENTER.
crh
RH : 25.21 Ref1 ?
RH : 25.22 Ref1 ?
RH : 25.22 Ref1 ? 30.1
Press any key when ready ...
RH : 25.19 Ref2 ?
OK
7.
OK
indicates that the adjustment has succeeded and the new
calibration coefficients are calculated and stored.
8.
Take the probe out of the reference conditions and replace the
filter.
Using MI70
To adjust the relative humidity:
1.
Connect MI70 to MMT162.
2. Turn
MI70
on.
3.
On the MI70, go to
Menu
→
Functions
→
Adjustments
Figure 14
Main Menu
Figure 15
Functions Menu