Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 31
Formatter string consists of quantities and modifiers. The MMT162
measures the following quantities:
-
water activity (aw)
-
temperature (T) (metric unit: ºC, non metric unit:ºF)
- ppm for transformer oil only (H
2
O)
When selecting the quantity, use the abbreviations of the quantities. The
modifiers are presented in Table 6 on page 31 below.
Table 6
Modifiers
Modifier Description
x.y
Length modifier (number of digits and decimal places)
#t Tabulator
#r Carriage-return
#n Line
feed
“” String
constant
#xxx
Special character, code "xxx" (decimal), for example
#027 for ESC
U5
Unit field and length (optional)
Example:
>form "aw=" 6.4 aw #t "t=" 6.2 t #r#n
1>
>send
aw= 0.2644
t= 25.50
>
>form "Oil ppm= " h2o " " u3 #r#n
>send
Oil ppm= 16.6 ppm
>
Command
FORM /
will return the default output format. The default
output format depends on the device configuration.
>form /
>send
aw= 0.087 T= 24.0 'C
>