Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 91
Modifier Description
"" String
constant
#xxx
Special character, code "xxx" (decimal), for example #027
for ESC
U5
Unit field and length
ADDR
Transmitter address with two characters [00...99]
ERR
Error flags for P, T, Ta, RH [0000 ... 1111], 0 = no error
STAT
Transmitter status in 7 character field, for example:
N 0
no heating
h 115
probe heating active, power 115/255
H 159.0
purge heating active, temperature 159 ºC
S 115.0
purge cooling active, temperature 115 ºC
X 95.0
sensor heating active, temperature 95 ºC
SN Transmitter
serial
number
TIME Time
[hh:mm:ss]
DATE Date
[yyyy-mm-dd]
OK
Pressure stability indicator, two characters [OK or " "]
CS2
Modulus-256 checksum of message sent so far, ascii
encoded hexadecimal notation
CS4
Modulus-65536 checksum of message sent so far, ascii
encoded hexadecimal notation
CSX
NMEA xor-checksum of message sent so far, ascii
encoded hexadecimal notation
A3H
Pressure tendency [* or 0...8]
Example:
>form "SSR=" 4.2 ssr U5 #t "Ts=" ts U3 #r #n
OK
>
>send
SSR= 22.51% Ts= 23.14'C
>
Command '
FORM /
' will return the default output format. The default
output format depends on the device configuration.
>form /
OK
>send
Tdf= 0.5 'C x= 4.0 g/kg SSR= 22.4 % Ts= 23.1 'C
>
UNIT
Use the command
UNIT
[
x
] to select metric or non-metric output units.
Syntax:
UNIT
[
x
]
where
M or N
M = metric units
x =
N =
nonmetric
units