Chapter 6 ___________________________________________________Calibration and Adjustment
VAISALA ______________________________________________________________________ 133
Example:
>ctext
Adjust. info : (not set) ? -60ºC
>
Use the command
CDATE
to input date to the adjustment information
field. Set the adjustment date in format YYYY-MM-DD.
Example:
>cdate
Adjust. date : (not set) ? 2004-05-21
>
Press the adjustment button on the motherboard inside the transmitter to
disable the adjustment function.
Two-Point Relative Humidity Adjustment
Using Display/Keypad
NOTE
Reference humidities must be ~0 %RH and 30 ... 75 %RH. A suitable
product for the two point RH calibration is for example the Vaisala
Humidity Calibrator HMK15.
1. Press
the
ADJ
button (see Figure 2 on page 23) to open the
ADJUSTMENT MENU
:
2. Select
Adjust T
d
measurement
,
press
►
button.
3. Select
2-point RH adjustment
, press
START
. At this point, the
device carries out the chemical purge.
4.
Remove the filter from the probe and insert the probe in the dry end
reference condition (~0 %RH). Press
OK
.
0706-018A
Figure 67
Inserting the Sensor in Reference Humidity 1