Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 121
Sets the RUN mode output interval. The time interval is used only when
the RUN mode is active.
Example:
The output interval is set to 10 minutes
>intv 10 min
Output intrv.
:
10 min
>
Setting RUN output interval to zero enables the fastest possible output
rate.
ADDR
Addresses are required only for POLL mode (see serial line command
on page 96). Use the
ADDR
[
aa
] command to input the RS-485
transmitter address.
where
aa
=
address (0 ... 99) (default = 0)
Example:
The transmitter is issued the address 99
>addr
Address : 2 ? 99
>
OPEN
[nn]
When all transmitters on the RS-485 bus are in POLL mode the
OPEN
[
nn
] command sets one transmitter temporarily to STOP mode so that
other commands can be issued.
Syntax:
OPEN
[
nn
]
where
nn
=
address of the transmitter (0 ... 99)
CLOSE
The
CLOSE
command switches the transmitter back to the POLL mode.