USER'S GUIDE____________________________________________________________________
140 _________________________________________________________________ M210762EN-D
Adjusting Analog Outputs
In the analog output calibration the analog output is forced to the
following values:
current output
= 2 mA and 18 mA
voltage output
= 10 % and 90 % of the range
Connect DMT346 to a calibrated current/voltage meter in order to
measure either current or voltage depending on the selected output type.
Use the display/keypad or the serial line to do this.
Adjusting Analog Outputs Using
Display/Keypad
1. Press
the
ADJ
button on the motherboard to open the
ADJUSTMENT MENU
.
2. Press
the
►
arrow button to select
Adjust analog outputs
.
3.
Select the output to be adjusted
Adjust analog output
1/2
, press
START
.
4.
Measure the first analog output value with a multimeter. Enter the
measured value by using the arrow buttons. Press
OK
.
5.
Measure the second analog output value with a multimeter. Enter
the measured value by using the arrow buttons. Press
OK
.
6. Press
OK
to return to the adjustment menu.
7. Press
EXIT
to close the adjustment and to return to the basic
display.