USER'S GUIDE____________________________________________________________________
110 _________________________________________________________________ M210762EN-D
Analog Output Tests
Use the display/keypad for testing to test the operation of the analog by
forcing the outputs to known values. Measure then the values with a
current/voltage meter.
Use the display/keypad for testing.
1. Open
the
MAIN MENU
by pressing any of the
▲▼◄►
arrow
buttons.
2. Select
System
by pressing the
►
arrow button.
3. Select
Diagnostics
by pressing the
►
arrow button.
4. Select
Analog output tests
by pressing the
►
arrow button.
5.
Select one of the testing options
Force 0 %/50%/100% of scale.
Press
TEST
to confirm your selection. All outputs are tested
simultaneously. The actual output value depends on the selected
range.
6. Press
OK
to stop testing. Press
EXIT
to return to the basic display.
ITEST
Use the serial line to test the operation of the analog outputs. Use the
command
ITEST
[
aa.aaa bb.bbb
] to force the analog outputs to entered
values. The set values remain valid until you issue the command
ITEST
without parameters or reset the transmitter.
Syntax:
ITEST
[
aa.aaa bb.bbb
]
where
aa.aaa
= Current or voltage value to be set for channel 1 (mA or V)
bb.bbb
= Current or voltage value to be set for channel 2 (mA or V)
Example:
>itest 20
Ch1 (Tdf ) : * 20.000 mA H'CCD3
Ch2 (x ) : * 20.000 mA H'CCCB
>itest 20 5
Ch1 (Tdf ) : * 20.000 mA H'CCD3
Ch2 (x ) : * 5.000 mA H'34BD
>