Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 87
Command Description
RESET
Reset the device
TEST Self-diagnostics
information
TIME Set
time.
UNIT
Display output units
VERS
Display the software version information
XPRES Set
pressure
(temporarily)
MOL/MOLI
View/set mole weight parameter
Getting Measurement Message from
Serial Line
Starting Continuous Outputting
R
Enter the
R
command to start the continuous output of measurements.
Example:
>r
Tdf=-20.6 'C H2O= 958 ppmV x= 0.6 g/kg
>r
Tdf=-20.7 'C H2O= 958 ppmV T= 23.8 'C RH= 3.3 %RH
>
If a value is too long to fit to the allocated space in the output, or if there
is an error in outputting the quantity, the value is displayed with stars '*'.
You can change the format of the output with the following commands:
- Outputting interval can be changed with the
INTV
command.
- Output message format can be changed with the
FORM
command
.
- Status of purge, sensor warming and AutoCal can be added with the
FST
command.
- Date and time information can be added with commands
FDATE
and
FTIME
.
Stopping Continuous Outputting
S
Use the
S
command to end the RUN mode. After this command all other
commands can be used. You can also press the
Esc
button or reset the
transmitter to stop the outputting.