Property
Description/Value
Without compensation (typical)
-0.08 % of reading/%O
2
Long-term stability
0 … 8 %CO
2
< ±0.2 %CO
2
/year
8 % … 12 %CO
2
< ±0.5 %CO
2
/year
12 % … 20 %CO
2
< ±1.0 %CO
2
/year
Table 72 Operating environment
Property
Description/Value
Operating temperature for CO
2
measurement
0 … +70 °C (+32 … +158 °F)
Max. temperature durability in standby mode
(sensor head only)
Max. +195 °C (+383 °F)
Heat sterilization 180 °C durability
At least 120 cycles
Storage temperature
-40 … +75 °C (-40 … +167 °F)
Pressure (compensated)
500 … 1100 hPa
Pressure (operating)
< 1500 hPa
Humidity
0 … 100 %, non-condensing
Condensation prevention
Sensor head heating when power on
Chemical tolerance
DMSO, IPA (70 %), H
2
O
2
(2000 ppm, non-
condensing), ethanol, acetic acid
Electromagnetic compatibility
EN61326-1, Generic Environment
Table 73 Inputs and outputs
Property
Description/Value
Digital outputs
I
2
C 5 V, RS-485 (2-wire with Vaisala Industrial
Protocol)
Analog output
0 … 20 mA (scalable)
max. load 600 Ω
Power consumption
< 1 W (pulsed)
Operating voltage
11 … 30 VDC
20 ... 30 VDC (with analog output)