4.11 Configuring analog output
Table 35 ASEL command
Syntax
Description
ASEL
[channel]
<cr>
Show the parameter and scaling of the analog
output in ppm.
channel
= Analog output channel, only 1 can
be selected.
ASEL
[channel] [parameter lowlimit
highlimit]
<cr>
Set the parameter and scaling of the analog
output.
•
parameter
= Parameter that is output on
analog channel. Only parameter available is
CO2.
•
lowlimit
= Lower limit of channel scaling in
ppm.
•
highlimit
= High limit of channel scaling in
ppm.
Example (change channel scaling to 0 … 5 %):
pass
1300
asel
1 co2 0 50000
Aout 1 quantity : CO2(0 ... 50000)
Table 36 AMODE command
Syntax
Description
AMODE
[channel]
<cr>
Show currently set analog output limits and
error level.
channel
= Analog output channel, only 1 can
be selected.
AMODE
[channel lo_value hi_value
error_value]
<cr>
Set new analog output limits and error output
value.
•
lo_value
= Low limit of the channel.
•
hi_value
= High limit of the channel.
•
error_value
= Error value of the channel.
GMP231 User Guide
M211501EN-E
44
Содержание CARBOCAP GMP231
Страница 1: ...M211501EN E User Guide Carbon dioxide probe for CO2 incubators GMP231...
Страница 5: ...Online store 87 Warranty 87 Technical support 87 Recycling 87 Table of contents 3...