6. Take a new clean filter and push it onto the sensor. Push and rotate the filter slightly to
make sure it is fully in place.
Figure 14 Changing the silicone plug
6.4 Calibration and adjustment
Calibrate and adjust the CO
2
measurement of GMP231 as needed. Before starting, read through
this section completely so that you are aware of your options, and the main factors that affect
the result.
Performing an accurate calibration and adjustment takes some time and
preparation. Instead of doing it yourself, contact Vaisala.
Calibration
means comparing GMP231 to a known reference, such as calibration
gas or a reference instrument. Correcting the reading of the GMP231 so that is
measures accurately is referred to as
adjustment
.
6.4.1 Calibration setup
GMP231 is designed to be calibrated while it remains installed through an incubator wall. If you
calibrate the GMP231 without it being installed, note the following:
Chapter 6 – Maintenance
71