Message Segment
Length
Content
Frame length
1 byte
Always 07
hex
.
Data
1 byte
Return code. See
Checksum
2 bytes
See
.
Table 65 Adjustment subcommands
Code
Adjustment operation
0
Start 1 point adjustment
1
Start 2 point adjustment
2
Record measured point 1, reference value is given in parameter
3
Record measured point 2, reference value is given in parameter
4
Cancel adjustment (previous adjustment is reverted)
5
End (adjustment is saved and new values are taken in use)
6
Revert to factory calibration (clear user adjustment)
Table 66 Adjustment parameters
Code
Purpose
0
All parameters (valid with revert to factory calibration)
1
CO
2
Table 67 Adjustment return codes
Code
Meaning
0
OK
1
Function not supported
2
Sequence error
3
Recorded-Reference difference too large
4
2-point adjustment: Points too close
5
Given reference value is invalid