Syntax
Description
TCMODE
[on | off | measured]
<cr>
Show current compensation value for
temperature compensation mode.
O2CMODE
[on | off]
<cr>
Show current compensation value for oxygen
compensation mode.
RHCMODE
[on | off]
<cr>
Show current compensation value for relative
humidity compensation mode.
•
on
= Compensation enabled using setpoint
value.
•
off
= Compensation disabled, neutral value
used.
•
measured
= Compensation enabled using
internal measurement. Available only for
pressure and temperature compensations.
Example (check temperature compensation mode):
pass
1300
tcmode
T COMP MODE : ON
Example (change temperature compensation mode to use internal measurement):
pass
1300
tcmode
measured
T COMP MODE : MEASURED
Table 23 ENV command
Syntax
Description
ENV
[temp|pres|oxy|hum] [value]
<cr>
Set new permanent compensation values and
store them in EEPROM.
•
temp
= Compensation temperature. Range
-40 ... 100 °C.
•
pres
= Compensation pressure. Range 500 ...
1150 hPa.
•
oxy
= Oxygen content of background gas.
Range 0 ... 100%.
•
hum
= Relative humidity of background gas.
Range 0 ... 100%.
GMP231 User Guide
M211501EN-E
36