• Probe detects a measurement malfunction. This means an actual measurement problem,
such as sensor damage or unsuitable environmental conditions.
• Measured value(s) are well outside the scaled output range.
The default error level depends on the output type:
• For 4 ... 20 mA output, the default error level is
3.6 mA
.
• For 0 ... 20 mA output, the default error level is
23 mA
.
The probe resumes normal operation of the analog output when the cause of the error state is
removed.
More information
‣
Analog output overrange behavior (page 15)
GMP231 User Guide
M211501EN-E