D291308.6r.fm
Verify Calibration Accuracy
Chapter 4: Maintenance & Calibration
75
Observe the graph results and readings to the right of the graph and below it.
7.
Let the chart record at least 50mm of travel then press
Stop Test
.
8.
The
Mean
value is displayed under the graph. It should be (+/-) 3g from the nominal value of
the weight used..
9.
If the mean value is out of spec, conduct
Calibration
and
Validation
again.
Figure 4.15
Figure 4.16
Содержание PT-55
Страница 2: ......
Страница 4: ......
Страница 76: ...66 PT 55 User s Guide Customizing the Main Screen Logo D291308 7M fm...
Страница 87: ...D291566 1 fm Appendix A Changing the Computer BIOS Setting A 1 Appendix A Changing the Computer BIOS Setting...
Страница 90: ......
Страница 94: ...PT 55 Spare Parts List D291427h fm Page 1 of 1...
Страница 106: ...Service and Parts Contacts 61053915 fm Page 2...
Страница 108: ......
Страница 110: ......
Страница 111: ......
Страница 112: ...751 Summit Avenue Mankato MN 56001 507 387 2039 FAX 507 387 2257 www vtekusa com Email info vtekusa com...