Calibration
Chapter 4: Maintenance & Calibration
69
Calibration
Equipment Required:
• 50 gram weight
• 100 gram weight
• 200 gram weight
Note:
Calibration weights sold separately.
Calibration Procedure
In the event that the load cell is replaced or if it is part of a maintenance routine, the load cell can
be recalibrated by following the steps outlined.
Note
: Once calibration has started, all calibration data will be lost and recalibrating is
required.
1.
From the
Main
screen, click the
Supervisor
button.
Figure 4.1
Figure 4.2