D291308.7M.fm
Main Screen Details
Chapter 3: Operation
29
Carrier Fields
.
The Carrier section tracks carrier tape information. Its fields include
Supplier, Type,
and
Lot
Number
Sealtape Fields
The
Sealtape
section is used to track information about the seal tape (also known as cover tape)
that is used in a test. Sealtape fields include
Supplier, Type,
and
Lot Number
.
Settings
The
Settings
section is used to configure the Peel Force
Test..
Note
: With the exception of
Pull Length
, the
Settings
fields may only be changed by using a supervisor pass-
word.
Pull Length
Setting the pull length allows the user to stop the peel force
at a desired length. For example, if the
Pull Length
is set at
150, the peel force test will stop at 150mm.
Pull Length
should be set between 0-200 mm.
Note
: If the
Pull Length
is set to 0, the pull test will default
to the
X Axis Maximum
which was set in the
Supervisor
Settings
screen
.
If the
X Axis Maximum
setting is different
than the
Pull Length
setting, the pull test will stop at whichever length is shortest.
Figure 3.3
Figure 3.4
Figure 3.5
Содержание PT-55
Страница 2: ......
Страница 4: ......
Страница 76: ...66 PT 55 User s Guide Customizing the Main Screen Logo D291308 7M fm...
Страница 87: ...D291566 1 fm Appendix A Changing the Computer BIOS Setting A 1 Appendix A Changing the Computer BIOS Setting...
Страница 90: ......
Страница 94: ...PT 55 Spare Parts List D291427h fm Page 1 of 1...
Страница 106: ...Service and Parts Contacts 61053915 fm Page 2...
Страница 108: ......
Страница 110: ......
Страница 111: ......
Страница 112: ...751 Summit Avenue Mankato MN 56001 507 387 2039 FAX 507 387 2257 www vtekusa com Email info vtekusa com...