56
PT-55 User’s Guide
Configuring & Running a Peel Test
D291308.7M.fm
7.
In
the
Main Screen
, type in the desired pull length
in the
Pull Length
field.
Pull Length
should
be set between 0-200 mm.
Note
: If the
Pull Length
is set to 0, the pull test will default to the
X Axis Maximum
which was
set in the
Supervisor Settings
screen
.
If the
X Axis Maximum
setting is different than the
Pull
Length
setting, the pull test will stop at whichever length is shortest.
8.
Click the
Save File
button
to create a
Master Form
, if desired.
Note
: for more details on creating a Master Form, proceed to the next section in this chapter.
9.
Ensure sealed tape sample is loaded properly and carriage is in the
Home
position.
Figure 3.36
Содержание PT-55