PT-55 User’s Guide
Calibration
D291308.6r.fm
Ensure the weight is suspended from the center of the calibra-
tion jig bearing and is stabilized (not swinging) to ensure accu-
rate calibration.
Note
: Hang weights over the calibration jig in the front of the carriage assembly before con-
necting them to the load cell clip.
Never
hang weights without using the calibration jig. Hang-
ing weights directly from the load cell may cause damage to the load cell.
8.
Repeat Step 7 with 100 gram and 200 gram weights, following the directions on the screen.
9.
When calibration is complete, the
Supervisor Settings
screen will once again be accessible
and a message will appear prompting to validate the calibration.