BioDoc-It
2
Imaging System
5
81-0358-01 Rev E
Specifications
Power Requirements:
100/115V, 50/60Hz; 3.1 Amps at 120 Volts
230V, 50/60Hz; 1.55 Amps at 230 Volts
Mains supply voltage fluctuations are not to exceed 10 percent of the nominal
supply voltage
Pollution Degree:
2
Installation Category:
II
Altitude:
Up to 2000m
Ambient Temperature:
5°C to 40°C
Humidity:
Maximum relative humidity of 80% for temperatures up to 31°C, decreasing
linearly to 50% maximum relative humidity at 40°C
Built-In Touch Screen Computer
Size
10.1“ touch screen
Screen
Multi-touch, 1280 x 800 Resolution
Operating System
Windows® 10 Home (32 bit)
Hard Drive Capacity
64 GB Flash Storage
Memory
2 GB RAM
Wireless Networking Capability
802.11 b/g/n Wi-Fi and Bluetooth 4.0
Connectivity Ports
1 USB (side of system)
6 USB (rear of system)
Software
VisionWorks
touch
Software
Cameras and Lenses
The BioDoc-It
2
315 is equipped with the GelCam 315 Camera, with a resolution of 5.0MP. The system is
equipped with an 8-48mm zoom lens, allowing the user to zoom in and out on the sample within the
darkroom.
The GelCam 315 uses a USB 2.0 connection. All camera settings are factory pre-set for optimum
performance when viewing gels and films under low light level conditions. Contact UVP Technical Support
before making any adjustments to camera settings.
Ethidium Bromide (EtBr) Emission Filter
The ethidium bromide (50mm
2
) UV-blocking bandpass interference filter blocks UV and IR radiation emitted
from the transilluminator. The filter is placed in the two-position emission filter assembly and allows
visualization of fluorophores from 580-630nm, targeting the ethidium bromide emission peak of 605nm.
Additional filters are available for other specific fluorophores, including custom filters. Filters can also be
removed when imaging non-fluorescent media (including protein gels, colony plates, etc.) in order to
produce brighter images. Contact UVP for ordering information.
Darkroom
Darkroom features include:
Integrated 10” touchscreen interface
Integrated transilluminator and epi (overhead) LED white lights
UV-safe gel viewer window built into the darkroom door
Side doors for easy access to the interior of the darkroom (for placement or excision of samples)
Two-position emission filter selector
UV safety interlock switch to disable UV transillumination when the main darkroom door is opened