
Upstream Instruction Manual
Released on 5/1/10 – R800001L
Page 29
UV Pure Technologies Inc. © 2010
S
ETUP
M
ENU
Setup Menu
NC 10-75
Language
English
Language
English/French
Remote StartStop
Disabled
Remote StartStop
Disable/Enable
Shutoff Valve
Not Installed
Shutoff Valve
Not Installed / Installed
Unit
Configuration
Enter Password
######
Remote Device
BIND #
Bind Remote
Device
Remote Device
BOUND
A
LARMS
&
T
ROUBLESHOOTING
M
ENU
Alarms and
Troubleshooting
Lamp Alarm -
Low UV Intensity
Lamps need to
burn in
Continue to
operate lamps
UV Air Temp.
High
Check if filter
dirty, fan fail
Check if room
temp within spec
UV Air Temp.
Low
Check if room
temp within spec
Switch lamps &
examine values
Only one lamp
may be low
Sensor may
have drifted
New sensor or
calibration reqd
Lamps require
replacement
Install 2 lamps
& reset counter
Front Panel ajar
Fully Close
front panel
Lamps not
striking
Front Panel ajar
Fully Close
front panel
Check lamps
or ballast
Lamp Lifetime
exceeded
Lamps require
replacement
Install 2 lamps
& reset counter
End of Lamp Life
Approaching
Lamp cycling
too often
Reduce the # of
lamp starts/day
ALARM
WARNING
ALARM
ALARM
WARNING
This column represents all possible
reasons for the fault.
Содержание UPSTREAM
Страница 47: ......
Страница 48: ...UV Pure Technologies Inc 60 Venture Drive Unit 5 Toronto Ontario M1B 3S4 www uvpure com ...