SU320CSX
4110-0446, Rev. B
Page 66 of 96
© 2016 UTC Aerospace Systems
Date Printed: 5-Dec-2016
This document does not contain any export controlled technical data.
Range
-50.00 to +70.00, 223.00 to 343.00 (K)
Type signed
decimal
Example
FPA:TEMP? -- query command
(one of)
37.81
-- return value
310.95 Kelvin -- return value
5.15.3.
Get Thermoelectric Cooler Lock Status
Description
Returns status of the thermoelectric cooler stabilization lock of the
focal plane array temperature to the setpoint. The temperature is
considered locked when the current temperature is within ±0.1°C of
the setpoint. The TEC should remain locked when the case
temperature is maintained within a given range, see Section 1.5.
Setting Type
N/A
Command TEC:LOCK?
Parameters none
Return Values
status
Range LOCKED
TEC
stabilized
NOT LOCKED TEC not stabilized
Type string
Example
TEC:LOCK? -- query command
LOCKED -- return value
5.15.4.
Get Thermoelectric Cooler Setpoint
Description
Returns the thermoelectric cooler temperature setpoint in degrees
Celsius.
Setting Type
Operational
Command TEC:SETPOINT?
Parameters none
Return Values
value
Range -20
to
80
Type integer
Example
TEC:SETPOINT? -- query command
22 -- return value