96
Reference
devices that open to atmosphere after reaching a certain
pressure.
Cal Value
: Cal Value = Pressure change during the leak test
step with a calibrated leak master – the Comp Value.
Calibration
: Comparison of a device (such as Sprint iQ) to a
standard that is in turn calibrated to an even more accurate
standard.
Calibration Data
: Values entered into Sprint iQ through
software calibration.
Calibration Menu
: Calibration allows comparison for
Sprint iQ to pressure or flow standards. The calibration
screen shows Sprint iQ’s actual reading and the pre-
programmed target value the technician compares to the
pressure or flow standard. Calibration should be
conducted only by qualified technicians who have proper
training and resources to calibrate Sprint.
Calibration Table
: Technicians don’t need to adjust
potentiometers, dipswitches, or other electrical
components to calibrate Sprint. All calibration is done by
making adjustments to Sprint iQ’s sensor linearization
tables.
Comp Value
: Pressure change during the leak test step with a
non-leaking master
Concurrent
: Testing multiple ports or multiple parts at the
same time is known as concurrent testing. If the tester has
multiple test ports and tests them all at the same time, it is
called a concurrent tester. See Sequential.
Contrast
: The contrast of the LCD can be adjusted for the
best viewing using the Display Contrast setting in the
Setup Menu.